MPD-120 - Fonte de Alimentação Chaveada 120Watts, Saída Dupla, Certificação Médica
Meanwell - MPD-120 - Fonte de Alimentação Chaveada 120Watts
MeanWell
Model: MPD-120
Product ID: 986
Product SKU: 986
New
In stock
/>

MPD-120 - Fonte de Alimentação Chaveada 120Watts, Saída Dupla, Certificação Médica
Estoque: 6
Marca:
MeanWellModelo:MPD-120
Referência:
MPD-120
Área: Alimentação Industrial
Característica: Sem Caixa
Certificações: Médica
Potência (W): 120
Saída: Dupla
AC input voltage | 90~264 VAC, 47~440Hz / 127~370VDC |
---|---|
AC input current | Maximum input current 2.9A at 115VAC, 60Hz or 1.7A at 230VAC, 60Hz with 100% output load. |
Inrush current | Less than 22A at 115VAC or 45A at 230VAC at cold start conditions |
Setup, rise time | 500ms, 20ms / 230VAC at full load, 1200ms, 50ms / 115VAC at full load |
Hold-up tiime | 80ms / 230VAC at full load,14ms / 115VAC at full load |
Leakage current | Leakage current is less than 180µA at 264VAC |
DC output adj. range | Between -5%~+10% rated output voltage |
Over load protection | Will be activated at 110~150% rated load. |
Over voltage protection | 115% ~ 135% rated output voltage |
Power good/Fail signal (optional) | At 90% of rated value, a+5V TTL signal will be sent out with a 10~500ms delay; at least |
1ms before the output voltage goes below 90% of the rated value,the TTL signal will be turned off. | |
Working temp. | operate from -20~70°C . |
Workinig humidity | 20~90% RH non-condensing. |
Storage temp., humidity | -40~+85 °C, 10~90% RH |
Temp. coefficient | ±0.04%/°C on all outputs at full load between 0~50°C of ambient temperature. |
Vibration | 2G of acceleration,vibrating frequency adjust from 10Hz ~500Hz within a 10 min cycles |
Safety standards | Medical:UL60601-1, TUV EN60601-1 approved |
Commercial: Also design refer to UL60950-1, TUV EN60950-1 | |
Withstand voltage | 4000VAC between input and output,1500VAC between input and F.G.,500VAC between output and F.G. |
Isolation resistance | >100M Ohms for I/P-O/P, I/P-FG, O/P-FG by using 500VDC test voltage. |
EMI compliance | Compliance Level EN55011, Class B,EN55022, Class B,EN61000-3-2,EN61000-3-3 |
EMS compliance | Compliance Level EN61000-4-2,Level 3, 8KV,EN61000-4-2 Level 2, 4KV |
EN61000-4-3 Level 2, 3V/m,Level 3, 10V/m,EN61000-4-4 Level 2, 1KV/5KHz,Level 3, 2KV/5KHz | |
EN61000-4-5,Level 4, 2KV/Line-Line 4KV/Line-Earth,EN61000-4-6 Level 2, 3Vrms/m,Level 3, 10Vrms/m | |
EN61000-4-8,Level 2, 3A/m,Level 3, 10A/m,EN61000-4-11 Compliance | |
ENV50204,Level 2, 3V/m, 900MHz,Level 3, 10A/m, 900MHz | |
MTBF | 262,100 hours min at full load and 25°C of ambient temperature. |
Características
- Área: Alimentação Industrial
- Característica: Sem Caixa
- Certificações: Médica
- Potência (W): 120
- Saída: Dupla